Crossover procedure for switching to a new batch of process control strips without misadjusting the processor.
When a roll of control strips (pre-exposed or PCE) is running out, a new roll must replace it. Because the emulsion of the new batch may give different densities, a crossover is essential; otherwise the processor may be adjusted incorrectly and film produced with low or high density.
Highlights
- Using the current control strips, confirm that the processor is in control and meets current aims, and correct it if necessary
- Splice three strips from the current batch and three strips from the new batch together and process them
- Measure the RD density of the pre-exposed strip or the aim step of the sensitometer exposure for the old and new strips
- Confirm that the old batch is within the allowed range
- Use the new RD or aim step value as the new control aim
- Repeat steps 1 to 4 for each processor
- Record the crossover and new densities on the process control chart or start a new chart for each processor
For more information, refer to D-17 Control Procedures for Source Document Microfilm Processing.
Source:Eastman Park Micrographics(epminc.com)(opens in a new tab)
Frequently asked questions
Why is crossover necessary when replacing a roll of control strips?
Because the emulsion of the new batch may give different densities, otherwise the processor may be adjusted incorrectly and produce film with low or high density.
How many control strips from the current and new batches are used for crossover?
Three strips from the current batch and three strips from the new batch are spliced together and processed.
What must be checked before setting a new control aim during crossover?
Using the current control strips, the processor must be confirmed to be in control and meet current aims, and corrected if necessary.
What values are measured to set the new control aim during crossover?
The RD density of the pre-exposed strip or the aim step of the sensitometer exposure for the old and new strips is measured, and the new RD or aim step value is used as the new control aim.
What must be recorded on the process control chart after crossover?
The crossover and new densities must be recorded on the process control chart or a new chart must be started for each processor.
Attachments
Comments
No comments yet. Be the first to share your view.
Leave a comment
More to read

Tech tip · Dental Imaging
Planmeca ProMax Cephalostat with ProTouch User's Manual
User's manual for the Planmeca ProMax Cephalostat with ProTouch touchscreen, for accurate head positioning and cephalometric imaging on 2D and 3D ProMax units.
Planmeca

Tech tip · Dental Imaging
ORTHOPANTOMOGRAPH OP 3D Pro User Manual
User manual and technical specifications for the OP 3D Pro digital dental imaging system with panoramic, cephalometric and Cone Beam 3D programs.
Instrumentarium Dental

Tech tip · Microfilm
Tech Tip 91: Choosing the Correct IMAGELINK Product
IMAGELINK camera, duplicating, Archive Writer and COM films and processing chemicals grouped by equipment and application, with catalog numbers.
Eastman Park Micrographics