Using a Hydrometer to Control Processor Chemistry

How to measure developer and fixer specific gravity with a hydrometer in deep tank processors, with EPM target values and a recommended hydrometer.

A hydrometer is an instrument for measuring Specific Gravity (the ratio of water to concentrated chemistry). Deep tank processors such as Allen or Houston Fearless use concentrated developer and fixer, and regular specific gravity checks ensure correct mixing and image quality.

A hydrometer is a hollow tube with a weighted bottom section and a graduated stem. In a denser liquid, less of the stem sinks below the surface. The reading is taken where the top of the meniscus meets the stem.

Highlights

  • Check specific gravity at least weekly
  • The test can be done in a graduated cylinder or directly in the chemical tank
  • The value must fall between the upper and lower limits set by the manufacturer
  • Too much water (weak mix) = low background density
  • Too much chemistry (strong mix) = high background density

Specific gravity specifications

Solution Mix ratio Specific Gravity
EPM Developer Concentrate 1:7 1.045 to 1.055
EPM Fixer Concentrate 1:3 1.080 to 1.105

Recommended hydrometer

Parameter Value
Supplier VWR International
Catalog Number 34610-002
Range 1.000 to 1.220
Approximate length 300mm

Source:Eastman Park Micrographics(epminc.com)(opens in a new tab)

Frequently asked questions

How is a hydrometer used to measure specific gravity in deep tank processors?

The hydrometer is placed in a graduated cylinder or directly in the chemical tank, and the reading is taken where the top of the meniscus meets the graduated stem.

What is the allowed specific gravity range for EPM Developer Concentrate at a 1:7 mix ratio?

The allowed range is 1.045 to 1.055.

What is the allowed specific gravity range for EPM Fixer Concentrate at a 1:3 mix ratio?

The allowed range is 1.080 to 1.105.

How does excess water affect background density in deep tank processor chemistry?

Too much water creates a weak mix and results in low background density.

Which hydrometer is recommended for measuring specific gravity in deep tank processors?

The recommended hydrometer is VWR International Catalog Number 34610-002, with a range of 1.000 to 1.220 and an approximate length of 300mm.

Attachments

Comments

No comments yet. Be the first to share your view.

Leave a comment

Answer with digits

More to read

  • ORTHOPANTOMOGRAPH OP 3D Pro User Manual

    Tech tip · Dental Imaging

    ORTHOPANTOMOGRAPH OP 3D Pro User Manual

    User manual and technical specifications for the OP 3D Pro digital dental imaging system with panoramic, cephalometric and Cone Beam 3D programs.

    Instrumentarium Dental

  • Tech Tip 91: Choosing the Correct IMAGELINK Product

    Tech tip · Microfilm

    Tech Tip 91: Choosing the Correct IMAGELINK Product

    IMAGELINK camera, duplicating, Archive Writer and COM films and processing chemicals grouped by equipment and application, with catalog numbers.

    Eastman Park Micrographics

Planning a new imaging room or upgrading equipment?

Tell us about your center. Our application specialists will recommend the right configuration and prepare a tailored proposal.