Skip to content
CPI Indico 100 RAD X-ray Generator - Image 1 of 2
CPI Indico 100 Generators

CPI Indico 100 RAD X-ray Generator

Model
Indico 100 RAD
Brand
CPI

Indico 100 RAD radiographic generator from 32 to 100 kW with output up to 400 kHz, kV/AEC and kV/mAs modes and one- or two-tube operation.

Description

Indico 100 RAD radiographic generator from 32 to 100 kW with output up to 400 kHz, kV/AEC and kV/mAs modes and one- or two-tube operation.

  • Dedicated console for radiographic (RAD) applications
  • Power levels of 32, 40, 50, 65, 80 and 100 kW with high frequency output up to 400 kHz
  • kV/AEC, kV/mAs and kV/mA/Time operating modes, with 150 kV standard
  • 1 or 2 tube operation, kV/mAs Stepping, multiple DR and AEC interfaces, Falling Load
  • 5 programmable image receptors

Indico 100 Touchscreen (RAD and RF)

  • 15 inch Touchscreen display with PowerPC-based CPU
  • More than 20,000 anatomical programs (APR) and technique selection with memory
  • Multilingual support, backup and restore, custom graphical interface
  • Adjustable viewing angle, wall mount and optional hand switch

GenWare service software

  • Generator programming through GenWare or the operator console
  • Very quick and easy installation
  • System diagnostics for remote analysis
  • Automatic X-ray tube calibration
  • Library of more than 300 X-ray tubes

Technical specifications

General
Control processor16 Bit CPU
StarterIntegrated Low Speed Starter or Dual-Speed Starter
Auxiliary cabinet120/240 VAC (optional)
Technique selectionkV/AEC, kV/mAs or kV/mA/ms
DocumentationManuals on CD
X-ray generator
RAD or RF power levels32/40/50/65/80/100 kW
Special Procedures power levels80/100 kW
Generator frequencyUp to 400 kHz
Inverter moduleFET Inverter Modules
GeneratorHigh Frequency Output (400 kHz max.)
kVp range and steps40 - 150 kV in 1 kV steps
kVp accuracy±(2% + 1 kV) for 70 - 85 kV, ±(5% + 1 kV) for 40 - 150 kV
High voltage ripple<1 kV @ 110 kV
Tube current32 kW: 10 - 400 mA; 40 kW: 10 - 500 mA; 50 kW: 10 - 630 mA; 65 kW: 10 - 800 mA; 80 and 100 kW: 10 - 1000 mA
mA accuracy±(5% + 1 mA) for exposures > 5 ms, ±20% for exposures ≤ 5 ms
Exposure time1 ms to 6300 ms in 1 ms steps
Exposure timer accuracy±(2% + 0.5 ms) from 5 to 6300 ms, ±(10% + 1 ms) for ≤ 5 ms
mAs range (non-AEC)32 and 40 kW: 0.1 - 630 mAs; 50 to 100 kW: 0.1 - 1000 mAs
mAs accuracy±(10% + 0.2 mAs) for exposures > 5 ms
Tube
Rotor supplyLow Speed Starter, Dual Speed Starter optional
Connectivity
Communication portsRS232, RS422
Power
Maximum voltage150 kV standard
Power supply1 Phase 230 VAC, 3 Phase 400 or 480 VAC
Power input, 32 and 40 kW models1 Φ/208-230 VAC or 3 Φ/400-480 VAC
Power input, 50 to 100 kW models3 Φ/400-480 VAC
Line voltage range±10%
Manual power derating<10% line
Dimensions & weight
Cabinet dimensions549 x 457 mm, height 1226 to 1480 mm depending on configuration

Accessories

  • Touchscreen console with wall mount (optional)
  • Hand switch (optional)
  • Auxiliary cabinet (optional)
  • DAP or Air Kerma display (optional)