Article1 min read
A-1630 Microfilm Process Control Sheet
Standard A-1630 form for daily logging of control strip densities, temperature, speed and chemical specific gravity in a microfilm processor.
The Conventional Process Control Sheet (A-1630) is a standard log sheet for daily monitoring of microfilm processor performance. With this form the operator records control strip density readings and machine operating conditions for each run and can quickly spot any deviation from the allowed range.
Highlights
- Records three density values: High Density (HD), Reference Density (RD) and Low Density / D-min (LD)
- The allowed range for the Aim Density is ±.12, and the chart is graduated to ±.20 for each value
- Room for three entries per day (a.m. / p.m.) for each date
- Columns for developer and water temperature, time, processor speed (Proc Speed) and Specific Gravity
- Records mix ratio and replenishment rate for developer and fixer in mL/ft or mL/min
- Records Dwell Time, processor number, Control Step and Emulsion Batch
Instructions for use
Dwell Time is determined by timing the film as it travels through the developer tank (or a tank of the same size), from the top roller of the rack to the next roller, while the processor is running. Specific gravity is measured with a hydrometer with a range of 1.0 to 1.22 while the recirculation pumps are turned off.
Source:Eastman Park Micrographics(epminc.com)(opens in a new tab)
Frequently asked questions
Which density values are recorded on form A-1630?
Three density values: High Density (HD), Reference Density (RD) and Low Density / D-min (LD).
What is the allowed range for Aim Density on form A-1630?
The allowed range is ±.12, and the chart is graduated to ±.20 for each value.
How many daily entries can be recorded for each date on form A-1630?
Three entries per day (a.m. / p.m.) for each date.
How is Dwell Time measured in the microfilm processor?
By timing the film as it travels through the developer tank (or a tank of the same size), from the top roller of the rack to the next roller, while the processor is running.
How and with what instrument is specific gravity measured?
With a hydrometer with a range of 1.0 to 1.22 while the recirculation pumps are turned off.
Which machine operating conditions are recorded on form A-1630?
Developer and water temperature, time, processor speed (Proc Speed), specific gravity, mix ratio and replenishment rate for developer and fixer, Dwell Time, processor number, Control Step and Emulsion Batch.
Attachments
Comments
No comments yet. Be the first to share your view.
Leave a comment
More to read

Article · Mammography
Agfa Mammography Portfolio
Agfa's vendor-neutral mammography portfolio: Mamoray film, DRYSTAR printers, MUSICA processing, CR Mammo and DR Mammo Retrofit.
Agfa

Article · DR Detectors
Agfa Digital Radiography (DR) Portfolio Overview
Overview of the Agfa DR portfolio: DR 600, DR 800, DX-D 300, DR 400, DX-D 100+, DR 100e, DR Retrofit and DR detectors, all powered by MUSICA image processing.
Agfa

Article · Veterinary
Agfa Equine Digital Radiography with MUSICA 3+ Image Processing
Agfa veterinary DR for equine and large animal imaging with MUSICA 3+ processing, wireless detectors and a low-cost upgrade path from film or CR.
Agfa